Description
- Calibrate all kind of single/three phase electronic/inductive active and reactive energy meter
- Automatically or single-step manually test basic error, creeping, starting, standard deviation of single/three phase energy meter, save user-defined calibrating scheme in Auto calibration
- Support variables influencetests, such as: Voltage, frequency, harmonics, reverse phase sequence, Voltage unbalance, etc
- Support special tests in energy meter type test, such as: phase triggering, pulse train triggering, etc
- Support DLT-645 communication protocol,multi-testing of multifunction meter such as: demand indication error, demand cycle error, time slot switching error, energy error of different tariff period, etc
- With OCXO, daily error of energy meter can be detected
- Output 2nd~63rd
- 8” TFT screen, interface friendly, easy to operate;
- Equipped with RS232, Ethernet, and WiFi interface,MTS-100 can either support stand-alone operation, PC control or handy wireless terminal control;
- Communicate with tested meters thru RS-485;
- Automatically calculate the error of meter, and the related data can be downloaded thru U-disk;
- Self-protection, alarming and displaying overload location for equipment output overload, Voltage short-circuit, Current open-circuit;
- Automatically detect, diagnose and alarm for failure.
- Remotely updating online, easily achieve software updating;
- Support calibration by user locally.
AC Voltage output:
Range: 100V, 220V, 380V
Adjusting range: (0~120) %RG, RG refers to range
Adjusting resolution: 0.01%RG, 0.1%RG, 1%RG, or 10% RG
Stability: 0.01%/2min (Class 0.05), 0.02%/2min (Class 0.1)
Distortion: ≤0.2% (Non capacitive load)
Max output load: 20VA each phase
Accuracy: 0.05%RG (Class 0.05), 0.1%RG (Class 0.1)
AC Current output:
Range: 0.05A, 0.2A, 1A, 5A, 20A, 100A
Adjusting range: (0~120) %RG, RG refers to range, similarly hereafter
Adjusting resolution: 0.01%RG, 0.1%RG, 1%RG, or 10% RG
Stability: 0.01%/2min (Class 0.05), 0.02%/2min (Class 0.1)
Distortion: ≤0.2% (Non capacitive load)
Max output load: 120VA (100A range)
Accuracy: 0.05%RG (Class 0.05); 0.1%RG (Class 0.1)
Power output
Power output stability: 0.01%/2min (Class 0.05), 0.02%/2min (Class 0.1)
Active/reactive power measurement accuracy: 0.05%RG;
Phase output
Output adjusting range: 0°~360°
Output adjusting resolution: 10°, 1°, 0.1°, or 0.01°
Resolution: 0.01°
Measurement accuracy: 0.05°
Power factor output:
Adjusting range: -1 ~ 0 ~ +1
Measuring resolution: 0.0001
Measurement accuracy: 0.0005
Frequency output
Adjusting range: 40Hz ~70Hz
Output adjusting resolution: 1Hz, 0.1Hz, 0.01Hz, or 0.001Hz
Resolution: 0.001Hz
Accuracy: 0.002 Hz.
Harmonic setting
Harmonic order: 2nd ~63rd
Harmonic amplitude: 0~40%
Harmonic angle: 0°~359.99°
Harmonic set error: 2~31times: ≤±0.1%, 32~63 times: ≤±0.2%;
Energy error measurement:
Active energy basic error limit:
±0.05%RD (Voltage 15V~660V, Current 0.02A~120A, PF≥0.5)
±0.1%RD (Voltage 15V~660V, Current 0.01A~0.02A, PF=1)
Reactive energy basic error limit:
±0.1%RD (Voltage 15V~660V, Current 0.02A~120A, PF≥0.5);
±0.2%RD (Voltage 15V~660V, Current 0.01A~0.02A, PF=1);
Energy pulse input
Input pulse: 3 channels active, 3 channels reactive
Other parameters:
Power supply: 90-265VAC/DC
Power frequency: 50Hz~60Hz
Power consumption: 50VA~1000VA
Environment condition: 20℃~30℃,Humidity: RH≤85%
Storage environment: -20℃~50℃
Size: 600mm(L)× 440mm(W)× 266mm(H)
Weight: 35kg
MTS-100 contains 2 models with different accuracy below:
MTS-100C: Accuracy Class 0.05
MTS-100B: Accuracy Class 0.1
Customers also viewed: